

BAUER Air Purification Systems purify high pressure air to a quality that meets or exceeds the requirements of CGA Pamphlet G-7, Compressed Air for Human Respiration, ANSI /CGA G-7.1, Commodity Specification for Air, Grade E, and all other recognized standards for breathing air.
| UP TO 6000 PSI SERVICE | |
| PURIFICATION AIR PROCESSING | SYSTEM CAPABILITY+ |
|---|---|
| P0++ | 3,200 cu. ft. |
| P2 | 40,000 cu. ft. |
| 2 SECURUS | 67,000 cu. ft. |
| P5 | 90,000 cu. ft. |
| 5 SECURUS | 150,000 cu. ft. |
| P10 | 140,000 cu. ft. |
| 0 SECURUS | 230,000 cu. ft. |
| 2 SECURUS* | 420,000 cu. ft. |
| 4 SECURUS* | 650,000 cu. ft. |
| P31++ | 11,760 cu. ft. |
| P41 | 28,700 cu. ft. |
| P41S | 47,000 cu. ft. |
| P42 | 64,000 cu. ft. |
| P42S | 107,000 cu. ft. |
| P43 | 100,000 cu. ft. |
| P43S | 164,000 cu. ft. |
* P12 and P14 only available with SECURUS.
+ Air Processing Capability based on standard inlet conditions, 68o F., a Pressure of 14.70 psia and a relative humidity of 36% (0.0750 density).
++ 5000 PSI Maximum Pressure, Triplex Design is not incorporated on a panel.
SECURUS II ensures 100% utilization of the air processing cartridges because the moisture sensor is embedded directly in the molecular sieve. The SECURUS cartridge, with its moisture sensor, is the last cartridge in the air processing system. SECURUS II is powered from the PLC Control System and provides for automatic warning and shutdown when the SECURUS cartridge becomes saturated. The SECURUS II can be mounted locally to the SECURUS cartridge chamber or to the compressor’s control panel. SECURUS II is available for working pressures from 2000 – 6000 PSIG and flow rates to 125 SCFM.